Evaluation and comparison of automated biopsy devices. Work in progress
C R Mladinich1, N Ackerman, C R Berry
1Department of Radiology, College of Medicine, University of Florida, Gainesville 32610-0374.
Radiology
|September 1, 1992
Summary
Four automated biopsy devices showed similar performance in liver and kidney sampling. The Bard Monopty device yielded heavier samples, but all devices were satisfactory with no significant differences in tissue preservation or organ injury.
Area of Science:
- Veterinary Medicine
- Medical Devices
- Surgical Technology
Background:
- Automated biopsy devices are increasingly used for tissue sampling in veterinary medicine.
- Ultrasound guidance enhances precision in organ biopsy procedures.
- Comparative studies are essential to evaluate the performance of different biopsy devices.
Purpose of the Study:
- To compare the performance of four automated biopsy devices (Bard Biopty, Bard Monopty, Microvasive ASAP 18, Medical Device Technologies Ultra-Cut) for liver and kidney sampling.
- To assess sample characteristics, tissue preservation, and organ trauma associated with each device.
Main Methods:
- Eight dogs underwent ultrasound-guided liver and kidney biopsies using four automated devices.
- A total of 96 liver and 96 kidney samples were collected.
- Sample length, weight, histological preservation, crush artifact, cellular integrity, and organ trauma were evaluated.
Main Results:
- No significant differences were observed in sample lengths or histological quality (preservation, crush artifact, glomeruli/lobules, portal triads) among the four devices.
- The Bard Monopty device produced significantly heavier kidney (30.8%) and liver (31.6%) samples compared to the others.
- Renal subcapsular hematomas were common, with no significant difference in renal trauma between devices. One instance of severe liver injury occurred.
Conclusions:
- All four automated biopsy devices are satisfactory for ultrasound-guided liver and kidney sampling in dogs.
- Device selection can be based on personal preference, as no significant differences in tissue quality or organ injury were found.
- The Bard Monopty device's tendency to yield heavier samples may be a consideration in device choice.