Measurement of CT section thickness by using the partial volume effect.

J Shiraishi1, K Tsuda, Y Inoue

  • 1Department of Radiology, Osaka City University Hospital, Japan.

Radiology
|September 1, 1992
PubMed
Summary

A new method accurately calculates computed tomographic (CT) section thickness using two materials. This technique offers improved accuracy over the aluminum ramp method, especially for thinner sections.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Endovascular Treatment for Infra-inguinal Autologous Saphenous Vein Graft Occlusion Using Self Expanding Nitinol Stents.

EJVES short reports·2017
Same author

Using breast radiographers' reports as a second opinion for radiologists' readings of microcalcifications in digital mammography.

The British journal of radiology·2014
Same author

Completion phenomenon in transcortical sensory aphasia.

Behavioural neurology·2014
Same author

Memory and cerebral blood flow in cases of transient global amnesia during and after the attack.

Behavioural neurology·2014
Same author

Is response rate increment obtained by molecular targeted agents related to survival benefit in the phase III trials of advanced cancer?

Annals of oncology : official journal of the European Society for Medical Oncology·2010
Same author

Effect of preservation solution on graft viability in single-lung transplantation from heart-beating donors in pigs.

Transplantation proceedings·2007

Area of Science:

  • Medical Imaging
  • Radiological Physics

Background:

  • Accurate measurement of computed tomographic (CT) section thickness is crucial for image quality and diagnostic accuracy.
  • Existing methods, such as the aluminum ramp method, have limitations, particularly with thin sections.

Purpose of the Study:

  • To develop and validate a novel method for calculating CT section thickness.
  • To compare the accuracy of the new method against the established aluminum ramp method.

Main Methods:

  • A method was developed to calculate section thickness based on mean CT numbers of two materials within a section.
  • A cylindrical water phantom with an acrylic rod was used for measurements.
  • Section thickness was measured using two different CT scanners.

Related Experiment Videos

Main Results:

  • The developed method showed good agreement with the aluminum ramp method for section thicknesses greater than 4 mm (approx. 10% difference).
  • For thinner sections (1-2 mm), the new method demonstrated significantly better accuracy, with differences of 30%-50% compared to the aluminum ramp method.
  • The new method's measurements were independent of the CT scanner's filter function, enhancing its reliability.

Conclusions:

  • The novel CT section thickness calculation method is accurate and reliable.
  • This method provides a more precise alternative to the aluminum ramp method, particularly for thin CT sections.
  • Its independence from filter functions contributes to its superior accuracy.