Related Experiment Video
Updated: Aug 24, 2026

Standardized Method for Measuring Collection Efficiency from Wipe-sampling of Trace Explosives
Published on: April 10, 2017
Implementation of CIEMAT/NIST LSC efficiency tracing method in KRISS: (204)T1 standardization
K B Lee1, Jong Man Lee, Tae Soon Park
1Korea Research Institute of Standards and Science, P.O. Box 102, Yuseong, Daejeon 305-600, South Korea. lee@kriss.re.kr
Abstract:
The CIEMAT/NIST efficiency tracing method has been widely used for radionuclide standardization. The method consists of the theoretical calculations of counting efficiencies and the experimental data obtained from a commercially available double photomultiplier detector. We have recently implemented the technique using (3)H as a tracer. Our implementation of the method was validated by participating into an international comparison of activity measurements of a solution of (204)T1. We report in this paper the procedure employed to implement the technique and compare our measurement result of (204)T1 activity with that of the international comparison.