Gain and maximum output of two electromagnetic middle ear implants: are real ear measurements helpful?
Ad Snik1, John Noten, Cor Cremers
1Department of Otorhinolaryngology, University Hospital Nijmegen, The Netherlands. a.snik@kno.umcn.nl
Abstract:
We compared the output of two electronic middle ear implants: the Otologics MET device and the Vibrant Soundbridge device. Both devices were programmed in the linear amplification mode. Aided minus unaided sound pressure levels recorded in the ear canal (objective gain) were compared to unaided minus aided soundfield thresholds (functional gain) in 13 patients with severe sensorineural hearing loss. In addition, input/output characteristics were studied with the help of ear canal measurements. Objective gain was consistently lower than functional gain, with wide variation between patients and frequencies. Using input/output data measured in the ear canal in combination with functional gain data, the mean maximum output of the two devices was estimated, expressed in dB SPL. In comparison to NAL-R target values, (functional) gain was adequate; however, the maximum output was low, especially for the Vibrant Soundbridge device.