Methods to determine drain volume for peritoneal dialysis clearances
Judith Bernardini1, Tracey Florio, Filitsa Bender
1University of Pittsburgh, Pittsburgh, Pennsylvania, USA. bernardi@pitt.edu
Summary
For peritoneal dialysis (PD) clearance studies, measuring drain volume by cycler for automated PD (APD) and by subtracting bag weight for continuous ambulatory PD (CAPD) is accurate and convenient.
Area of Science:
- Nephrology
- Dialysis Technology
Background:
- Accurate measurement of peritoneal dialysis (PD) effluent volume is crucial for assessing dialysis adequacy.
- Traditional methods for measuring drain volume can be cumbersome and pose risks of fluid exposure.
Purpose of the Study:
- To compare the accuracy and convenience of three distinct methods for measuring drain volume in PD clearance studies.
- To identify the most practical and precise method for both automated PD (APD) and continuous ambulatory PD (CAPD).
Main Methods:
- A prospective comparison of drain volume measurement techniques was conducted.
- Three methods were evaluated: graduated cylinder, weight-based measurement (with and without bag weight), and cycler-indicated volume.
- The study included 28 PD patients (15 CAPD, 28 APD) across two dialysis clinics.
Main Results:
- No statistically significant differences in measured volumes were found among the three methods for either APD or CAPD.
- Effluent volume measurement was found to be more accurate when the empty bag weight was subtracted in the weight-based method.
- The cycler-measured volume for APD and the weight-based method (subtracting bag weight) for CAPD demonstrated high accuracy.
Conclusions:
- For APD, using the cycler's measurement of initial drain, total ultrafiltration, and prescribed inflow volume is a precise and convenient method.
- For CAPD, weighing drained bags and subtracting the empty bag weight provides an accurate measurement.
- Both recommended methods minimize staff exposure to body fluids, enhancing safety in PD clearance studies.