Related Experiment Video
Updated: Aug 24, 2026

Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
Forehead is as sensitive as finger pulse oximetry during general anesthesia
Shigekazu Sugino1, Noriaki Kanaya, Masahito Mizuuchi
1Department of Anesthesiology, Sapporo Medical University School of Medicine, Sapporo, Japan.
Purpose:
To compare the performance of a forehead probe to a conventional finger pulse oximetry probe in anesthetized patients.
Methods:
Eighteen patients participated in the study. Each probe was connected to a Nellcor N-550 pulse oximeter. Anesthesia was induced and maintained with propofol. After intubation, the patients received air to achieve a steady-state of peripheral arterial oxygen saturation (SpO(2)). Ventilation was interrupted to induce a hypoxic state. As soon as one of the two SpO(2)'s decreased to 90%, the patients' lungs were ventilated with 100% oxygen. To evaluate the performance of the two pulse oximeters, time to the lowest (TL), time of recovery (TR) and lag times to beginning of SpO(2) decrease (Lag) were measured.
Results:
There were no significant differences in TL and TR between forehead and finger pulse oximetry under normal perfusion conditions during general anesthesia. When the axillary artery was compressed to mimic reduced peripheral perfusion, SpO(2) in the forehead decreased sooner than in the finger during hypoxia. The forehead and finger TLs were similar, however, TR was significantly longer in the finger.
Conclusion:
The forehead SpO(2) sensor can be used as an alternative to the conventional finger sensor during general anesthesia.
More Related Videos
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
06:04Objective Nociceptive Assessment in Ventilated ICU Patients: A Feasibility Study Using Pupillometry and the Nociceptive Flexion Reflex
Published on: July 4, 2018
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Stages of General Anesthesia
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...