Redo surgery after permanent pacemaker implantation: performance indicator of surgical techniques and pacing
1Department of Cardiology, Tata Main Hospital, Jamshedpur. vbharat@satyam.net.in
Background:
The incidence of permanent pacemaker-related complications is reducing due to advancement of technology and increasing operator experience. There are only few series from India reporting the annual complication rates from a single center over the years. This is a series of 782 pacemakers implanted over 20 years in a secondary healthcare set-up.
Methods And Results:
Eighty-two patients underwent redo surgery, either for a procedure-related complication (n=34) or for replacement of a malfunctioning/end-of-life pacemaker (n=48). Through critical analysis and corrective measures, all the procedural complications were reduced to less than 4% of the annual implantations. The introduction of a pacemaker follow-up clinic contributed to reducing the rate of elective replacement for battery depletion from 19.17% of the implanted VVI pacemakers to 0.63%.
Conclusions:
Despite being a low-volume center, with less than 100 pacemakers implanted annually, the performance of our pacing practice has shown continuous improvement.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiomyopathy VII: Pre and Post Operative Nursing Management