Does device-based testing save time during automatic implantable cardioverter-defibrillator implantation?

Anoop K Gupta1, Atul Khasnis, Ranjan K Thakur

  • 1Thoracic and Cardiovascular Institute, Michigan State University and Sparrow Health System, Lansing, MI, USA. anoopgupta@msn.com

Indian Heart Journal
|May 8, 2004
PubMed
Summary

External vs. device-based defibrillation testing showed similar outcomes. Neither method significantly reduced procedure time or improved defibrillation parameters, suggesting comparable efficacy for implantable cardioverter-defibrillator implantation.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same authorSame journal

Indian society of electrocardiology position statement on improving CPR awareness in the Indian public.

Indian heart journal·2026
Same author

Establishing a national cardiopulmonary resuscitation program in Indian high schools and colleges: An expert collaborative call to action and implementation model.

Heart rhythm·2026
Same author

Accessory Pathway Associated With a Venous Lake of the Right Atrial Thebesian Vein.

JACC. Clinical electrophysiology·2026
Same author

Recurrent ventricular tachycardia associated with non-compliance with thyroxine treatment for hypothyroidism.

BMJ case reports·2026
Same author

Multicentric reticulohistiocytosis.

Cleveland Clinic journal of medicine·2026
Same author

Establishing a National Cardiopulmonary Resuscitation Program in Indian High schools and colleges: An expert collaborative call to action and implementation model.

Indian pacing and electrophysiology journal·2025

Area of Science:

  • Cardiology
  • Biomedical Engineering

Background:

  • Defibrillation testing for implantable cardioverter-defibrillators (ICDs) can utilize external devices or device-based testing during implantation.
  • The comparative advantages of these two testing methodologies remain insufficiently studied.

Purpose of the Study:

  • To compare the efficacy and procedural characteristics of external cardiac defibrillation versus device-based testing during ICD implantation.

Main Methods:

  • A randomized trial involving 74 patients (72% male) comparing external cardiac defibrillation and device-based testing groups.
  • Key parameters assessed included R-wave, pacing threshold, pacing impedance, defibrillation threshold, defibrillation pathway impedance, and total procedure time.

Main Results:

Related Experiment Videos

  • No significant differences were observed in R-wave, pacing threshold, pacing impedance, defibrillation threshold, defibrillation pathway impedance, or total procedure time between the two groups.
  • Lead repositioning or replacement was necessary in three patients within the external cardiac defibrillation group.
  • Conclusions:

    • Device-based defibrillation testing did not significantly decrease procedure time compared to external testing.
    • Both testing methods yielded similar lead and defibrillation parameters, indicating comparable effectiveness.