Related Experiment Video
Updated: Aug 24, 2026

Mechanical Ventilation Boot Camp Curriculum
Published on: March 12, 2018
A closed-loop mechanical ventilation controller with explicit objective functions
Frederico C Jandre1, Alexandre V Pino, Ivanir Lacorte
1Biomedical Engineering Program/COPPE, Federal University of Rio de Janeiro, RJ 21945-970, Brazil. jandre@peb.ufrj.br
Abstract:
A closed-loop lung ventilation controller was designed, aiming to: 1) track a desired end-tidal CO2 pressure (Pet CO2), 2) find the positive end-expiratory pressure (PEEP) of minimum estimated respiratory system elastance (Ers,e), and 3) follow objective functions conjectured to reduce lung injury. After numerical simulations, tests were performed in six paralyzed piglets. Respiratory mechanics parameters were estimated by the recursive least squares (RLS) method. The controller incorporated a modified PI controller for Pet CO2 and a gradient descent method for PEEP. In each animal, three automated PEEP control runs were performed, as well as a manual PEEP titration of Ers,e and a multiple PetCO2 step change trial. Overall performance indexes were obtained from PEEP control, such as minimum Ers,e (37.0 +/- 4.5 cmH2O x L(-1)), time to reach the minimum Ers,e (235 +/- 182 s) and associated PEEP (6.5 +/- 1.0 cmH2O), and from Pet CO2 control, such as rise time (53 +/- 22 s), absolute overshoot/undershoot of PetCO2 (3 +/- 1 mmHg), and settling time (145 +/- 72 s). The resulting CO2 controller dynamics approximate physiological responses, and results from PEEP control were similar to those obtained by manual titration. Multiple dependencies linking the involved variables are discussed. The present controller can help to implement and evaluate objective functions that meet clinical goals.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal and...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...