Factors influencing fetal pulse oximetry performance
Related Experiment Video
Updated: Aug 14, 2026

Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
1OB Scientific, Inc., Germantown, WI 53022, USA. jim.reuss@obscientific.com
Abstract:
Fetal pulse oximetry shares many problems with traditional pulse oximetry, but also poses unique challenges that may compromise accurate SpO2 determination and data availability. The purpose of this review is to characterize the factors that may influence fetal pulse oximetry and their relationship to sensor design. A review of the literature of pulse oximetry identifies the factors that have been shown to influence fetal pulse oximetry performance, as well as other factors from traditional pulse oximetry that may also be expected to have an effect. Sensor design choices, including monitoring site, wavelength, and configuration, are related to the incidence and mitigation of these factors. Sensor designs may be characterized by monitoring site, means of retention, and operating mode (reflectance versus transmission). The factors influencing fetal pulse oximetry may be categorized as follows: fetal physiology, tissue characteristics at the monitoring site, sensor-tissue interface, and external influences. Monitoring site selection is of paramount importance in reducing the impact of interfering factors on fetal pulse oximetry performance. Many factors of importance in traditional pulse oximetry have yet to be characterized as far as their potential for interference in fetal pulse oximetry.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Special considerations while measuring pulse
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...