Measuring grip strength in normal adults: reference ranges and a comparison of electronic and hydraulic instruments
Nicola Massy-Westropp1, Wayne Rankin, Michael Ahern
1Department of Orthopaedic Surgery, Daw Park, SA, Australia.
Purpose:
To determine reference ranges for peak, average, and final adult grip strength over 10 seconds by using an electronic dynamometer, and to compare results from hydraulic and electronic dynamometers.
Methods:
The hand-grip strengths of 476 healthy adult subjects were tested using the electronic (Grippit; AB Detektor, Goteborg, Sweden) and hydraulic (Jamar; Smith and Nephew, Memphis, TN) dynamometers.
Results:
Age- and gender-specific reference ranges for the Jamar and Grippit dynamometers are presented. Bland-Altman analysis of the differences between the results obtained using the 2 instruments revealed a bias (mean difference) of 22 N (Jamar - Grippit) and limits of agreement of -86 to 129 N (mean +/- 2 SD), which indicates that grip measurements may vary by up to 215 N between instruments.
Conclusions:
The study yielded population reference ranges of peak, average, and final strength over a 10-second grip assessment using an electronic dynamometer. Results from the Grippit and Jamar dynamometers are similar; however, the dynamometers cannot be interchanged. The Grippit provides information about endurance and fatigue of grip over 10 seconds, showing differences between right- and left-dominant adults.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Non-destructive Tests for Concrete Strength