X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
The dynamic range of digital radiographic systems: dose reduction or risk of overexposure?
W E R Berkhout1, D A Beuger, G C H Sanderink
1Department of Oral and Maxillofacial Radiology, Academic Centre for Dentistry Amsterdam (ACTA), Amsterdam, The Netherlands. e.berkhout@acta.nl
Objectives:
To investigate the range of diagnostically acceptable digital radiographs and film as a function of exposure time, as well as the relationship to dose reduction and consequences for dental practice.
Methods:
Five systems for intraoral radiography were used to take a series of radiographs, with increasing exposure times, of five different dry bone specimens. Seven observers evaluated the 25 series of radiographs. The observers had to determine which radiographs of each series were acceptable for dental diagnostics and which radiograph of each series they preferred.
Results:
For Ektaspeed Plus film, the exposure time for the preferred radiograph was 0.52 s, with a range of diagnostically acceptable radiographs from 0.23-1.02 s. The preferred radiograph of the solid-state systems required less radiation than film (Sirona, 0.13 s; MPDx 0.35 s). The exposure range of these systems is narrow. In contrast, the exposure range of the phosphor plate systems is very wide. The preferred radiograph of the phosphor plate systems required high exposure (Digora, 1.21 s; Gendex DenOptix, 1.16 s).
Conclusions:
All digital systems require less exposure than film for diagnostically acceptable radiographs, but this is less obvious for preferred radiographs. Solid-state systems alert the dentist when a too long exposure time is used by a lack of image quality; phosphor plate systems, however, produce good quality radiographs even at high exposure times, which may result in an unnecessarily high dose.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Pharmacokinetic–Pharmacodynamic Relationship: Exposure, Response and Effect