Orbital floor reconstruction with poly-L/D-lactide implants: clinical, radiological and immunohistochemical study in
R Kontio1, R Suuronen, Y T Konttinen
1Department of Oral and Maxillofacial Surgery, Helsinki University Hospital, P.O. Box 263, 00029 HUCH, Finland. risto.kontio@hus.fi
Abstract:
In this study the reconstruction capacity of orbital wall in sheep was evaluated when poly-L/D-lactide (PLDLA96) implants were used for large blow-out defects in 18 sheep. The contralateral side, where the defects healed spontaneously, served as controls. The follow-up was 12, 16, 22 and 36 weeks. Healing was evaluated clinically, radiologically, histologically and immunohistochemically. Physiochemical properties of the implants were also studied. At first, the implants were surrounded by elastic capsules, which gradually ossified. At 36 weeks, 60% were still visible and deformed but surrounded by bone. Light microscopy revealed a low grade inflammatory reaction. Expression of Tn-c and cFn was intense throughout the study. Shear strength decreased gradually and was not measurable after 16 weeks. Crystallinity increased steadily from 1.5 to 29.30% and molecular weight decreased from 49,000 to 4186. In CT, the final bony defect was smaller in the reconstructed sides than in the controls. Based on this study it can be concluded that PLDLA96 implant provokes a local inflammation, which does not prevent bone healing. The deformation of the implant, however, indicates that this PLDLA96 plate is not suitable for orbital floor reconstruction.