Related Experiment Video
Updated: Aug 16, 2026

Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
High-salt stacking principles and sweeping: comments and contrasts on mechanisms for high-sensitivity analysis in
1Department of Physiology and Biophysics, University of California, Medical Sciences I, C-337, Irvine, CA 92697, USA. jfpalmer@uci.edu
Abstract:
High-salt stacking in electrokinetic chromatography (EKC) is defined and contrasted to the sweeping method. A recent paper argued the two methods are identical, where high concentrations of micelle in the sample were intended to mimic the effect of high-salt stacking. However, high micelle concentration in the sample matrix in EKC is analogous to using a high-conductivity sample instead of a low-conductivity sample in field amplified stacking. High-salt stacking does not require a sample free of pseuostationary phase, only a sample with a high-mobility co-ion compared to the separation buffer electrokinetic vector. High-salt stacking uses a discontinuous buffer system and should not be confused with continuous buffer stacking systems such as sweeping.
Related Concept Videos
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...