Related Experiment Video
Updated: Aug 24, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
A thermoluminescent dosimetry postal dose inter-comparison of radiation therapy centres in Malaysia
P Rassiah1, K H Ng, L A DeWerd
1Clinical Oncology Unit, University of Malaya Medical Centre, Kuala Lumpur, Malaysia. rassiah@uthscsa.edu
Abstract:
A thermoluminescent dosimetry (TLD) postal dose inter-comparison was carried out amongst radiotherapy centres in Malaysia. The aim of this TLD inter-comparison was to compare the uniformity involved in the measurement of absorbed dose among the participating centres. A set of 5 TLD chips placed within acrylic trays were mailed to all participating centres for irradiation to an absorbed dose to water of 2 Gy. Measurements were made for 6 MV and 60Co photon beams. Results show an agreement of +/- 5% for all but three radiotherapy centres. The ratios of the TLD readings to that of the reference centre are comparable with other national/regional dose inter-comparisons. The importance of a proper ongoing quality assurance program is essential in maintaining the consistency and uniformity of doses delivered.