Related Experiment Video
Updated: Aug 24, 2026

Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
Percutaneous retrieval of 20 centrally dislodged Port-A catheter fragments
Juhn-Cherng Liu1, Hsiuo-Shan Tseng, Chia-Yuen Chen
1Department of Radiology, Taipei Veterans General Hospital, Taipei, Taiwan.
Abstract:
A total of 20 patients with centrally embolized Port-A catheter fragments underwent the percutaneous retrieval procedures at our hospital. The causes of the dislodgement of these catheters included bad connection between the port and catheter, angulation or distortion at the anastomosis site, severing the catheter during insertion and removal of the catheter, improper catheter position and fatigue of the catheter. Consequently, improper procedure handling by the inexperienced surgeons could be considered as the most frequent cause of embolization. The percutaneous retrieval procedures were successful in all centrally embolized fragments. Of 20 successful retrievals, 16 were performed will loop snare catheters and 4 with Dormia basket retrievers.