Pump-induced haemolysis: a comparison of short-term ventricular assist devices

M Bennett1, S Horton, C Thuys

  • 1Cardiac Surgical Unit, Royal Children's Hospital, Parkville, Victoria, Australia. martin.bennett@rch.org.au

Perfusion
|May 28, 2004
PubMed

Insights

Centrifugal pumps, like the Rotaflow, show less blood damage (hemolysis) than roller pumps during prolonged use. The Rotaflow is suitable for short-term ventricular assist devices.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Devices
  • Hemodynamics

Background:

  • Centrifugal pumps are increasingly used in medical devices.
  • Pump-induced hemolysis is a critical factor in device performance and patient safety.
  • Roller pumps have historically been used but may cause significant hemolysis.

Purpose of the Study:

  • To compare the hemolytic properties of a standard roller pump with two centrifugal pumps (Biomedicus BP 50 and Jostra Rotaflow).
  • To evaluate the suitability of the Jostra Rotaflow for short-term ventricular assist applications.

Main Methods:

  • An in vitro test circuit was established for each pump type.
  • Circuits were operated for a six-day period, with experiments repeated five times.
  • Plasma hemoglobin levels were measured to quantify hemolysis.

Main Results:

  • The standard roller pump exhibited significantly higher hemolysis compared to both the Biomedicus BP 50 (p = 0.022) and the Jostra Rotaflow.
  • A statistically significant difference in hemolysis was observed between the Biomedicus BP 50 and the Jostra Rotaflow on day six (p = 0.016).
  • The Jostra Rotaflow demonstrated lower hemolysis levels than the Biomedicus BP 50.

Conclusions:

  • Newer generation centrifugal pumps, such as the Jostra Rotaflow, offer reduced hemolysis compared to traditional roller pumps.
  • The Jostra Rotaflow is a viable and potentially superior option for short-term ventricular assist due to its lower hemolytic profile.