Clinical versus instrumented knee testing on autopsy specimens
B C Fleming1, R J Johnson, E Shapiro
1McClure Musculoskeletal Research Center, Department of Orthopaedics and Rehabilitation, University of Vermont, Burlington 05405.
Clinical Orthopaedics and Related Research
|September 1, 1992
Summary
The Genucom and Knee Signature System (KSS) show inconsistent results for evaluating knee laxity. A subjective clinical examination is more reliable than these instrumented devices for assessing anterior cruciate ligament injuries.
Area of Science:
- Orthopedics
- Biomechanical Engineering
- Sports Medicine
Background:
- Accurate assessment of anteroposterior knee laxity is crucial for diagnosing anterior cruciate ligament (ACL) injuries.
- Instrumented devices like the Genucom Knee Analysis System and Knee Signature System (KSS) aim to provide objective laxity measurements.
- Previous studies have explored the utility of these devices, but their clinical reliability remains a subject of investigation.
Purpose of the Study:
- To compare the accuracy, reliability, and reproducibility of the Genucom and KSS for anteroposterior knee laxity evaluation.
- To compare the performance of these instrumented devices against a subjective clinical examination.
Main Methods:
- Specimens with and without ACL deficiency were tested using the Genucom and KSS to compare relative laxity changes.
- A clinical study involved ten patients undergoing Lachman and drawer tests performed by two independent examiners using the Genucom, KSS, and subjective clinical assessment.
Main Results:
- Both devices detected relative laxity changes but differed in absolute values.
- The Genucom exhibited interexaminer variability during Lachman testing.
- The subjective clinical examination demonstrated higher reliability than both the Genucom and KSS for both Lachman and drawer tests.
- High intraexaminer variability was observed for both Genucom (±4.2 mm) and KSS (±2.8 mm) at 95% confidence limits.
Conclusions:
- The Genucom and KSS demonstrate significant variability, questioning their accuracy and reproducibility in clinical settings.
- Subjective clinical examination is currently more reliable for assessing knee laxity compared to the evaluated instrumented devices.
- Further research is needed to improve the reliability and accuracy of instrumented knee laxity assessment systems.