Peripheral arterial tonometry events and electroencephalographic arousals in children

Riva Tauman1, Louise M O'Brien, Benjamin T Mast

  • 1Kosair Children's Hospital Research Institute and Division of Pediatric Sleep Medicine, Department of Pediatrics, University of Louisville, Louisville, KY 40202, USA.

Sleep
|May 29, 2004
PubMed

Insights

Peripheral arterial tonometry (PAT) detects sympathetic activity during sleep in children. While PAT events correlate with arousals, they often occur without electroencephalographic arousals, limiting its use for measuring pediatric sleep arousals.

Area of Science:

  • Pediatric Sleep Medicine
  • Autonomic Nervous System Function
  • Cardiorespiratory Physiology

Background:

  • Peripheral arterial tonometry (PAT) measures sympathetic nervous system activity and arousals in adults.
  • Its utility in identifying sleep arousals in children, particularly those with sleep-disordered breathing, requires investigation.

Purpose of the Study:

  • To assess the association between PAT events and electroencephalographic (EEG) arousals in healthy children and those with sleep-disordered breathing (SDB).
  • To determine the sensitivity and specificity of PAT for detecting arousals in pediatric populations.

Main Methods:

  • A prospective cohort study involving 60 children (20 with obstructive sleep apnea syndrome, 20 with mild SDB, 20 controls) aged 5.7-16.5 years.
  • Polysomnography with simultaneous PAT recording was performed.
  • PAT events were defined by signal attenuation, and various PAT indexes were calculated.

Main Results:

  • Total PAT index significantly correlated with total and spontaneous arousal indexes, particularly in children with obstructive sleep apnea syndrome (r=0.71, P<.0001).
  • PAT demonstrated high sensitivity (95%) but low specificity (35%) for identifying EEG arousals.
  • PAT detected pathologic arousal indexes with good accuracy (AUC=0.79, P=.002).
  • 92% of respiratory events were associated with PAT attenuations, compared to only 35% with EEG arousals.

Conclusions:

  • Sleep arousals in children involve increased sympathetic discharge, reflected by PAT signal changes.
  • A substantial number of PAT attenuations occurred without concurrent EEG arousals, suggesting autonomic arousals not captured by standard criteria.
  • PAT cannot currently replace EEG for measuring arousals in children, and adult arousal criteria may be insufficient for pediatric populations.
Abstract