Related Experiment Video
Updated: Aug 24, 2026

Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Real-time three-dimensional echocardiography for rheumatic mitral valve stenosis evaluation: an accurate and novel
José Zamorano1, Pedro Cordeiro, Lissa Sugeng
1Echocardiography Laboratory, Hospital Clínico San Carlos, Madrid, Spain. jlzamorano@vodafone.es
Objectives:
Our aim was to assess which echo-Doppler method has the best agreement with the mitral valve area (MVA) invasively evaluated by the Gorlin's formula. We also evaluated the feasibility and reproducibility of real-time three-dimensional echocardiography (RT3D) for the estimation of MVA and the Wilkins score in patients with rheumatic mitral stenosis (RMVS).
Background:
Real-time three-dimensional echocardiography is a novel technique that allows us to visualize the mitral valvular anatomy in any desired plane orientation. The usefulness and accuracy of this technique for evaluating RMVS has not been established.
Methods:
We studied a series of consecutive patients with RMVS from two tertiary care hospitals. Mitral valvular area was determined by conventional echo-Doppler methods and by RT3D, and their results were compared with those obtained invasively. Real-time three-dimensional echocardiography planimetry and mitral score were measured by two independent observers and then repeated by one of them.
Results:
Eighty patients with RMVS comprised our study group (76 women; 50.6 +/- 13.9 years). Compared with all other echo-Doppler methods, RT3D had the best agreement with the invasively determined MVA (average difference between both methods and limits of agreement: 0.08 cm(2) [-0.48 to 0.6]). Interobserver variability was as good for RT3D (intraclass correlation coefficient [ICC] = 0.90) as for pressure half-time (PHT) (ICC = 0.95). For PHT and RT3D, the intraobserver variability was similar (ICC 0.92 and 0.96, respectively). Real-time three-dimensional echocardiography valvular score evaluation showed a better interobserver agreement with RT3D than with 2D echocardiography.
Conclusions:
Real-time three-dimensional echocardiography is a feasible, accurate, and highly reproducible technique for assessing MVA in patients with RMVS. Real-time three-dimensional echocardiography has the best agreement with invasive methods.
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management