A simplified radiographic method for measuring bone-component motion in total knees.

P S Walker1, S Sathasivam

  • 1Institute of Orthopaedics, University College London, Stanmore, U.K.

Journal of Biomechanics
|September 1, 1992
PubMed
Summary

A new radiographic method using standard X-rays and computer analysis allows for routine assessment of implant stability and bone integration. This technique enhances accuracy for multi-center trials, improving prediction of long-term outcomes.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Neuromuscular scoliosis surgery in children.

BJA education·2025
Same author

Feasibility of single-use 3D-printed instruments for total knee arthroplasty.

The bone & joint journal·2019
Same author

Gallium Phosphide photoanode coated with TiO<sub>2</sub> and CoO<sub>x</sub> for stable photoelectrochemical water oxidation.

Optics express·2019
Same author

Frontotemporal lobar degeneration with motor neurone disease (FTLD/MND): Presentations in psychiatric practice.

International journal of psychiatry in clinical practice·2014
Same author

Treatment of posttraumatic elbows with a new hinge-distractor.

Orthopedics·2014
Same author

Analysis of an early intervention tibial component for medial osteoarthritis.

Journal of biomechanical engineering·2014

Area of Science:

  • Orthopedic biomechanics
  • Medical imaging analysis
  • Implant stability assessment

Background:

  • Measuring relative motion between bone implants and surrounding bone is crucial for evaluating component stability and predicting outcomes.
  • The established RSA (Roentgen Stereophotogrammetric Analysis) method offers high accuracy (0.1 mm) but requires specialized equipment and software.
  • A need exists for a more accessible method for routine implant assessment and multi-center clinical trials.

Purpose of the Study:

  • To develop and validate a novel method for measuring implant-bone motion using standard radiographs.
  • To enhance the accuracy of component stability assessment for routine clinical use and multi-center studies.
  • To improve the prediction of potential future outcomes for orthopedic implants.

Main Methods:

Related Experiment Videos

  • Utilized standard anteroposterior (A-P) and mediolateral (M-L) radiographs.
  • Developed computer software to estimate and correct for out-of-plane rotations relative to the radiographic film.
  • Calculated bead heights relative to the component, enhancing accuracy in positional measurements.

Main Results:

  • Theoretical errors in measurement were demonstrated to be less than 0.04 mm within the expected range of rotations.
  • Experimental validation using simulated bones showed 95% confidence limits for axial displacement under +/- 0.3 mm.
  • Rotational accuracy in the plane of the radiograph was found to be within 0.6 degrees.

Conclusions:

  • The developed method provides a viable, more accessible alternative to RSA for assessing implant stability.
  • This technique is suitable for studying knee components, where subsidence can range from 0.5-2 mm over two years.
  • The enhanced accuracy supports its use in routine assessments and multi-center trials for improved comparative analysis.