A better method of quality control for 99mTc-tetrofosmin
Richard Hammes1, Lori A Joas, Thomas E Kirschling
1Nuclear Medicine E1/382C, University of Wisconsin Hospital and Clinics, Madison, WI 53706, USA. rj.hammes@hosp.wisc.edu
Journal of Nuclear Medicine Technology
|June 4, 2004
Summary
A new silica cartridge method offers a faster, more accurate quality control for technetium-99m-tetrofosmin radiochemical purity. This convenient technique surpasses traditional methods, improving diagnostic accuracy in nuclear medicine.
Area of Science:
- Nuclear Medicine
- Radiopharmaceutical Quality Control
- Analytical Chemistry
Background:
- The standard quality control (QC) for technetium-99m-tetrofosmin requires time-consuming, meticulous techniques for accurate radiochemical purity (RCP) assessment.
- Existing alternative methods have limited validation, often only with pertechnetate, which may not reflect actual impurities in (99m)Tc-tetrofosmin preparations.
- Accurate RCP is crucial for ensuring the efficacy and safety of radiopharmaceuticals in diagnostic imaging.
Purpose of the Study:
- To develop and validate a convenient, accurate, and rapid new method for assessing the radiochemical purity of (99m)Tc-tetrofosmin using silica cartridges.
- To compare the performance of the novel silica cartridge method against the traditional package insert technique and instant thin-layer chromatography (ITLC).
Main Methods:
- A novel quality control method utilizing silica cartridges with a 70:30 methanol:water solvent system was developed.
- The method was validated against various technetium-99m radiopharmaceuticals, including sodium pertechnetate ((99m)TcO(4)), (99m)Tc-glucoheptonate, (99m)Tc-sulfur colloid, and low-purity (99m)Tc-tetrofosmin.
- Validation involved comparing results with the package insert method and ITLC across multiple nuclear pharmacies, assessing factors like sample volume and flow rate.
Main Results:
- The silica cartridge method demonstrated robustness, unaffected by sample volumes (25-100 µL), unlike ITLC.
- A critical flow rate of ≤5 mL/min was identified for the silica cartridge method, which exhibited significantly lower variability (SD and variance) compared to ITLC.
- The silica cartridge method showed near-perfect correlation (R=0.99) with the gold standard package insert technique, achieving 100% sensitivity, accuracy, and specificity in identifying batches with <90% RCP.
Conclusions:
- The developed silica cartridge method is a superior alternative for (99m)Tc-tetrofosmin radiochemical purity testing, scoring 90.2 compared to 65.8 for ITLC based on established criteria.
- The study demonstrated that common validation impurities like free (99m)TcO(4) are not the significant radiochemical impurities in (99m)Tc-tetrofosmin, invalidating prior alternative methods.
- This new method is fast, convenient, accurate, safe, and economical, though precise control of flow rate is essential for optimal performance.