Related Experiment Video
Updated: Aug 24, 2026

Synthesis and Operation of Fluorescent-core Microcavities for Refractometric Sensing
Published on: March 13, 2013
Ultracompact biochemical sensor built with two-dimensional photonic crystal microcavity
E Chow1, A Grot, L W Mirkarimi
1Agilent Technologies, Inc., 3500 Deer Creek Road, Polo Alto, California 94304, USA. edmond_chow@agilent.com
Abstract:
We report an experimental demonstration of an ultracompact biochemical sensor based on a two-dimensional photonic crystal microcavity. The microcavity, fabricated on a silicon-on-insulator substrate, is designed to have a resonant wavelength (lambda) near 1.5 microm. The transmission spectrum of the sensor is measured with different ambient refractive indices ranging from n = 1.0 to n = 1.5. From observation of the shift in resonant wavelength, a change in ambient refractive index of delta(n) = 0.002 is readily apparent. The correspondence between absolute refractive index and resonant wavelength agrees with numerical calculation to within 4% accuracy. The evaporation of water in a 5% glycerol mixture is also used to demonstrate the capability for in situ time-resolved sensing.