Comparison of the Roche LightCycler vanA/vanB detection assay and culture for detection of vancomycin-resistant enterococci from perianal swabs
Related Experiment Video
Updated: Jul 24, 2026

A 1.5 Hour Procedure for Identification of Enterococcus Species Directly from Blood Cultures
Published on: February 10, 2011
L M Sloan1, J R Uhl, E A Vetter
1Division of Clinical Microbiology, Mayo Clinic and Foundation, 200 First St., S.W., Hilton 470, Rochester, MN 55905, USA.
Abstract:
We compared the performance characteristics of a real-time PCR method, the LightCycler vanA/vanB detection assay (Roche Diagnostics Corporation, Indianapolis, Ind.) to that of Enterococcosel agar (BBL, Sparks, Md.) for direct detection of vancomycin-resistant enterococci (VRE) from 894 perianal stool swabs. For 421 of 894 swabs, the result for LightCycler PCR was compared to an Enterococcosel plate containing vancomycin at 6 microg/ml; for the remaining 473 swabs, the result for LightCycler PCR was compared to an Enterococcosel plate containing 8 microg/ml vancomycin. The LightCycler method produced considerably more positive results than either the Enterococcosel plate containing vancomycin at 6 microg/ml (n = 25 versus n = 11; sensitivity, 100%; specificity, 97%; positive predictive value [PPV], 42%; negative predictive value [NPV], 100%) or the Enterococcosel plate containing vancomycin at 8 microg/ml (n = 31 versus n = 10; sensitivity, 100%; specificity, 95%; PPV, 32%; NPV, 100%). When possible, additional testing, including culture, LightCycler PCR, and/or a conventional PCR method (PCR-restriction fragment length polymorphism assay), were performed on either the original specimens or original cultures or subsequent specimens for cases in which the original specimen was positive by LightCycler PCR but the Enterococcosel plate was negative. This additional testing demonstrated positive results for 7 of 14 (50%) evaluable discordant specimens which initially tested as LightCycler PCR positive but culture negative using the Enterococcosel plate containing vancomycin at 6 microg/ml and 12 of 17 (71%) evaluable discordant specimens which initially tested as LightCycler positive but culture negative using the Enterococcosel plate containing vancomycin at (8 microg/ml). These results demonstrate that the LightCycler VRE detection assay is considerably more sensitive than the standard culture method for detecting VRE directly from perianal swab specimens. The LightCycler assay also provides results much faster than culture (approximately 3.5 versus > or =72 h). The use of this test could have important implications for the effective control and prevention of nosocomial outbreaks of VRE.
More Related Videos
08:30One-day Workflow Scheme for Bacterial Pathogen Detection and Antimicrobial Resistance Testing from Blood Cultures
Published on: July 9, 2012
09:07Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
Related Concept Videos
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...
Urine Studies II: Urine Culture and Sensitivity Test
Automated Microbial Diagnostics