Related Experiment Video
Updated: Aug 24, 2026

Clinical Anthropometrics and Body Composition from 3-Dimensional Optical Imaging
Published on: June 7, 2024
Dual energy X-ray absorptiometry measurements in small subjects: conditions affecting clinical measurements
Winston W K Koo1, Elaine M Hockman, Mouhanad Hammami
1Department of Pediatrics, University of Tennessee, Memphis, USA. wkoo@wayne.edu
Objective:
To document the clinical and experimental situations that may affect DXA measurements in small subjects.
Methods:
49 piglets (886g to 21100g) had measurements with either of two pencil beam densitometers (QDR 1000W and QDR 2000 Plus, Hologic Inc, Waltham, MA) using commercial infant (IWB) and adult whole body (AWB) software v5.71p and v5.71 respectively. AWB scans were analyzed with three additional software versions. 35 infants (2115 to 11564g) had IWB measurements.
Results:
DXA measurements of total weight, bone mineral content, bone area, bone mineral density, fat and lean mass from IWB scans (all piglets) and from AWB scans (piglets >12 kg) were highly reproducible (p < 0.001). A statistically significant change occurred in at least one of the DXA measurements from the use of different platforms, variations in the amount and placement of covering (e.g., blanket), placement of the external calibration standard, presence of radiographic contrast material, presence of movement artifact, delivery of an intravenous fluid bolus prior to scanning or improper delineation of external calibration standard during analysis. Additionally, results varied amongst different versions of software as well as between IWB and AWB softwares.
Conclusion:
In small subjects, consistency in the DXA techniques is paramount for valid and meaningful comparison of DXA data in bone mass and body composition.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...