Determination of guaifenesin in human serum by capillary gas chromatography and electron capture detection
Maged H M Sharaf1, Dwight D Stiff
1Department of Pharmacognosy and Natural Products Chemistry, Faculty of Pharmacy, Al-Azhar University, Cairo, Egypt. magedsharaf@yahoo.com
Abstract:
A method for the quantitation of guaifenesin in human serum has been developed and validated. The procedure involves liquid-liquid extraction of the serum sample in the presence of mephenesin as an internal standard, followed by derivatization and analysis using capillary gas chromatography (GC) and electron capture detection (ECD). Different solvents were tested for extraction of guaifenesin from serum. n-Hexane/dichloromethane (1:1, v/v) gave the highest recovery and the lowest background and was chosen as the extraction solvent. After extraction, the residue of guaifenesin was derivatized at 60 degrees C for 30 min, with trifluoroacetic acid anhydride (TFAA) in toluene in the presence of pyridine. Excess trifluoroacetic acid anhydride was removed using dilute solution of ammonium hydroxide. The method proved to be linear over the range of 25.0-1000 ng/ml. Recovery of guaifenesin from spiked samples was consistent, averaging 75.5% at 50.0 ng/ml with a range of 72.0-80.0% (N = 8 determinations) and averaging 78% at 800 ng/ml with a range of 76.0-81.0% (N = 8 determinations). The internal standard recovery was also consistent averaging 72.8% with a range of 67.0-76.0% (N = 16 determinations).
More Related Videos
07:54Capillary Electrophoresis Mass Spectrometry Approaches for Characterization of the Protein and Metabolite Corona Acquired by Nanomaterials
Published on: October 27, 2020
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II