Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapy
Berry M van Gelder1, Frank A Bracke, Albert Meijer
1Department of Cardiology, Catharina Hospital, Eindhoven, The Netherlands. carlgr@cze.nl
The American Journal of Cardiology
|June 15, 2004
Summary
Optimizing the interventricular pacing interval (VV interval) in biventricular pacing significantly enhances left ventricular (LV) function in heart failure patients. This precise timing, measured by LV dP/dt(max), offers further improvements beyond standard simultaneous pacing.
Area of Science:
- Cardiology
- Biomedical Engineering
- Heart Failure Management
Background:
- Simultaneous biventricular pacing is known to improve left ventricular (LV) function in heart failure patients with LV asynchrony.
- The precise timing of the interventricular pacing interval (VV interval) may offer further optimization of LV function.
Purpose of the Study:
- To investigate the acute hemodynamic response to varying the VV interval.
- To assess the impact of VV interval optimization on LV function, using maximum LV dP/dt (LV dP/dt(max)) as the primary parameter.
Main Methods:
- A biventricular pacemaker was implanted in 53 patients with advanced heart failure (NYHA class III-IV), left bundle branch block, LV asynchrony, and QRS interval >150 ms.
- Optimization of atrioventricular and VV intervals was guided by LV dP/dt(max) measurements using a 0.014-in sensor-tipped pressure guidewire.
Main Results:
- Simultaneous biventricular pacing improved LV dP/dt(max) by 17% in ischemic cardiomyopathy, 18% in idiopathic dilated cardiomyopathy, and 21% in atrial fibrillation.
- Optimizing the VV interval yielded further absolute increases in LV dP/dt(max) of 8%, 7%, and 3% in the respective groups.
- Optimal VV intervals varied, with LV pacing first being the most common configuration for achieving maximum dP/dt(max).
Conclusions:
- Optimization of the VV interval significantly enhances LV dP/dt(max) compared to standard simultaneous biventricular pacing.
- The 0.014-in sensor-tipped pressure guidewire provides an easy, accurate, and reliable method for evaluating VV interval optimization.
- Tailoring the VV interval is a crucial step in maximizing hemodynamic benefits in patients receiving biventricular pacing for heart failure.