Digital mammography: quality and dose control.

Cosimo Di Maggio1, Mauro Gambaccini, Gisella Gennaro

  • 1Sezione di Senologia, Dipartimento Scienze Oncologiche e Chirurgiche, Università degli Studi, Padova, Italy.

La Radiologia Medica
|June 15, 2004
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A python framework for single-image characterization of X-ray focal spot distribution and detector point spread function.

Medical physics·2026
Same author

Adapting mammography DRLs for improved usability.

Radiation protection dosimetry·2026
Same author

Clinical Performance of Contrast-Enhanced Mammography Versus Breast MRI in Women at Increased Breast Cancer Risk.

Cancers·2026
Same author

Corrigendum: Phase-contrast breast CT: the effect of propagation distance (2018<i>Phys. Med. Biol</i>.<b>63</b>24NT03).

Physics in medicine and biology·2026
Same author

Corrigendum to "Stability of radiomic features in magnetic resonance imaging of the female pelvis: A multicentre phantom study". [Phys. Med. 130 (2025) 104895].

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2026
Same author

Radiomic approach to support multidisciplinary tumor board decision-making in locally advanced non-small cell lung cancer.

Frontiers in oncology·2026

Digital mammography equipment shows linear response and good reproducibility (<5%) in quality control tests. Careful parameter selection is crucial for accurate breast imaging, especially with varying thickness measurements.

Area of Science:

  • Medical Imaging Physics
  • Radiological Technology
  • Quality Assurance in Healthcare

Background:

  • Digital mammography requires robust quality control (QC) for accurate diagnosis.
  • Standardized QC parameters are essential for consistent performance evaluation.
  • Collaborative efforts between radiologists and physicists enhance QC protocols.

Purpose of the Study:

  • To define significant quality control parameters for digital mammography equipment.
  • To compare experimental data from multiple digital mammography units.
  • To establish reference values for digital mammography performance.

Main Methods:

  • Evaluated linearity, uniformity, reproducibility, and AEC stability on 5 digital mammography systems.
  • Utilized a simple test object for measurements.

Related Experiment Videos

  • Pre-checked X-ray beam quality and tube yield to isolate system response uncertainties.
  • Main Results:

    • All equipment demonstrated linear response and reproducibility errors below 5%.
    • Uniformity and grey-level compensation were within acceptable limits (5%).
    • Automatic Exposure Control (AEC) parameter selection varied, influenced by breast thickness measurement tolerances.

    Conclusions:

    • Results provide valuable reference data for users of specific digital mammography models.
    • Findings offer guidance for users of different digital mammography technologies, emphasizing the need for specific study.
    • Developing reliable QC protocols requires extensive investigation, cautioning against premature theoretical standards.