Contrast Imaging in Mouse Embryos Using High-frequency Ultrasound
Published on: March 4, 2015
In vitro analysis of ultrasound second generation contrast agent diluted in saline solution
Alfredo Goddi1, Raffaele Novario, Fabio Tanzi
1SME, Diagnostica per Immagini, Varese, Italy. goddi.alfredo@libero.it
Purpose:
To evaluate the changes in response of a second-generation sonographic contrast agent diluted in saline solution at different concentrations when different scanning techniques, and saline solution temperatures and pH were used.
Materials And Methods:
A series of tests was devised to analyse the behaviour of the sonographic contrast agent at different concentrations, temperatures, pH and scanning techniques. Latex balloons were used as phantoms. These were filled with 0.9% Sodium Chloride solution mixed with a suspension of sulphur hexafluoride microbubbles stabilised with phospholipids (SonoVue ) and scanned with the harmonic imaging technique at low acoustic pressure. For each image obtained, we calculated the mean grey-scale level values and the standard deviations of grey-scale level were calculated within a ROI on the US images obtained; the resulting data were used to create echogenicity curves of UCA the echo-enhancer over time at the different conditions tested.
Results:
We noted that SonoVue maintains adequate backscatter properties even at low concentration (0.15%) in N/S solution. The amount of echogenicity can be considered adequate for the average duration of an ultrasound examination. These properties are not affected by the scanning technique used on the phantom containing the UCA contrast agent/NaCl solution. The pH of the saline solution does not affect the survival of the microbubbles. Temperature becomes a significant value beyondover 33-37 degrees C; this condition favours rupture of the microbubbles with subsequent reduction of the echogenicity after 10-15 min, and complete disappearance after 30 min. US beam attenuation related to the concentration of the microbubbles in the saline solution volume has a non-linear behaviour; at low amplification levels, attenuation becomes more significant when using SonoVue concentrations higher than 0.4%. Unexpected greater improvement in US beam transmission was experienced for each of the three SonoVue concentrations tested, and especially for 0.2% and 0.4%, compared to saline solution alone.
Conclusions:
Our results show that a second-generation contrast agent may be used even diluted at low concentration in saline solution. This characteristic opens up new diagnostic perspectives possibilities for the use of contrast-enhanced harmonic imaging; in particular, its can be suggested used in hollow organs and functional studies. These in vitro results require by confirmation by clinical applications which are under evaluation and experimentation.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
06:02Multi-timescale Microscopy Methods for the Characterization of Fluorescently-labeled Microbubbles for Ultrasound-Triggered Drug Release
Published on: June 12, 2021
Related Concept Videos
Imaging Studies II: Ultrasonography
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Imaging Studies III: Computed Tomography