Related Experiment Video
Updated: Aug 23, 2026

Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
[Continuous Svo2 measurements and co-oximetry are not interchangeable immediately after cardiopulmonary bypass]
Karim Bendjelid1, Nicolas Schutz, Peter M Suter
1De la Division des Soins Intensifs de Chirurgie, Hôpitaux Universitaires de Genève, Genève, Suisse. Karim.Bendjelid@hcuge.ch
Objective:
To determine the accuracy of continuous (in vivo) measurement of mixed venous oxygen saturation (SvO(2)), using a fibreoptic catheter, in patients having had cardiopulmonary bypass (CPB).
Methods:
Using a pulmonary arterial catheter, we prospectively studied 14 patients (age 64 +/- 8) having had cardiopulmonary bypass. Mean hematocrit was 30 +/- 4%. The catheter was calibrated in vitro and in vivo, according to the manufacturer's instructions. Fifty-six simultaneous measurements of continuous SvO(2) (CSvO(2)) and measured SvO(2) (MSvO(2)) were taken with a co-oxymeter and the paired values were analyzed by the linear regression method. To make the two sets of measurements interchangeable, we established, a priori, a maximum limit of 3% (approximately 5% of the measurement), as being an acceptable difference between the two types of measurements.
Results:
All the measurements were obtained within four hours of the placement of the catheter. CSvO(2) was weakly correlated with MSvO(2), with a correlation coefficient of r(2) = 0.49 (P < 0.001). The Bland-Altman analysis demonstrates an objective mean bias of 0.8 +/- 3%, with 36% of the values measured falling outside clinically acceptable limits. For values of CSvO(2)
Conclusions:
Our results demonstrate that, in the immediate aftermath of cardiopulmonary bypass, the continuous measurement of SvO(2) and the co-oxymetry determination of mixed blood is not inter-changeable. Other studies will be necessary to confirm these results and enable us to understand the pathophysiological phenomena accounting for the lack of agreement between the two methods.
More Related Videos
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Cardiopulmonary Resuscitation II: ACLS Airway Management
Special considerations while measuring pulse