Accuracy of home blood pressure readings: monitors and operators
Trina Stryker1, Merne Wilson, Thomas W Wilson
1Department of Medicine and Cardiovascular Risk Factor Reduction Unit University of Saskatchewan, Saskatoon, Canada.
Objective:
To evaluate the accuracy of automated digital blood pressure monitoring devices and operators in the community. Also, we tested the effects of a simple education program, and looked for arm-arm differences.
Design:
Subjects who had bought their own automated digital blood pressure monitor were recruited via an advertisement in the local newspaper. On arrival, they were asked to record their blood pressure exactly as they would at home. The investigator noted any technique deficiencies then corrected them. Blood pressures were then recorded by the investigator and the subject, on opposite arms, simultaneously, and repeated with the arms switched. Finally, subjects recorded their blood pressure again. The subjects' readings were compared to the average of monitor and mercury readings using Bland-Altman methods.
Results:
A total of 80 subjects were tested. Before educating, subjects' systolic blood pressure (SBP) readings were +5.8+/-6.4 (standard deviation) mmHg greater than the mean of all readings, and diastolic blood pressure (DBP) were +1.3+/-4.0 mmHg; after educating they were +1.3+/-4.0 and -1.3+/-2.7 respectively. The monitors, as a group, were accurate, and met British Hypertension Society and AAMI highest standards. We found no differences among monitors that had been validated (n=26) and those that had not. There were differences between the arms: 5.3+/-5.2 mmHg for SBP and 3.4+/-3.3 mmHg for DBP. Most patients had never been informed by anyone of proper blood pressure measuring techniques.
Conclusions:
We conclude that home blood pressure measurement, as practiced in our community, is prone to error, mostly due to mistakes by the operator. These can easily be corrected, so that readings become more accurate. Attention should be paid to arm-arm differences.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)