Is SPECT/CT with a hybrid camera useful to improve scintigraphic imaging interpretation?
Orazio Schillaci1, Roberta Danieli, Carlo Manni
1Department of Biopathology and Diagnostic Imaging, University Tor Vergata, Rome, Italy. oschil@tiscalinet.it
Objectives:
To assess the usefulness of a hybrid imaging system (single-photon emission computed tomography/computed tomography, SPECT/CT) for functional anatomical mapping (FAM) using various radiotracers and the additional value of fused SPECT/CT images compared with SPECT alone.
Methods:
Eighty-one consecutive patients studied for various clinical situations were evaluated: 10 with 111In-pentetreotide, four with 123I-meta-iodobenzylguanidine (I-MIBG), five with 99mTc-labelled red blood cells, two with 99mTc-antigranulocyte antibodies, four with 99mTc-nanocolloids, 10 with 67Ga, seven with 99mTc-methylene diphosphonate (99mTc-MDP), nine with 99mTc-sestamibi, 21 with 99mTc-tetrofosmin, two with 201Tl, three with 99mTc-ethylcysteinate dimer (ECD), one with 99mTc-hexamethylpropyleneamine oxime (99mTc-HMPAO) and three with 123I-N-w-fluoropropyl-2-beta-carbomethoxy-3-beta-(4-iodophenyl) nortropanel (FPCIT). The acquisition of both anatomical (CT) and functional (SPECT) data was performed during a single session. SPECT data were first interpreted alone and then re-evaluated with the addition of FAM.
Results:
Transmission anatomical maps allowed for the precise anatomical localization of SPECT images in 79 of the 81 cases. SPECT/CT had a significant impact on the results in 33 (40.7%) of the 81 patients: FAM provided the correct localization of SPECT findings in 23 cases, allowed the definition of the functional significance of CT lesions in two and enabled the exclusion of disease in sites of physiological tracer uptake in eight.
Conclusions:
SPECT/CT with this hybrid system is a feasible technique yielding co-registered dual-modality images. FAM allows a more precise interpretation of scintigraphic studies in several cases and fused images can improve the diagnostic accuracy of SPECT in various clinical situations.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography