New blood volume monitoring method for hemodialysis: A-V pressure gradient measurement by synchronized one-point

Masami Oda1, Sanehiro Hokama, Kimio Sugaya

  • 1Department of Urology, School of Medicine, University of the Ryukyus, Okinawa, Japan. oda@med.u-ryuku.ac.jp

Artificial Organs
|June 24, 2004
PubMed

During hemodialysis, rapid ultrafiltration often causes symptomatic hypotension. To predict the occurrence of volume-dependent hypotension as early as possible, continuous hematocrit monitoring with the Crit-Line noninvasive monitor has been widely used to measure blood volume changes during hemodialysis. As another potential method of monitoring blood volume variations, we studied blood viscosity, which is theoretically associated with the pressure gradient across the dialyzer. Blood viscosity (calculated by the Hugen-Poiseuille formula) is a major determinant of the blood flow rate and is associated with the pressure difference between the postpump arterial (A) and venous (V) pressures. The A-V pressure gradient fluctuates due to pump pulsation, so we minimized this noise by always reading the pressure gradient at the same point out of 1400 partitions on the rotary pump. To test this synchronized one-point reading method, the A-V pressure gradient was measured using 3 different xanthan gum solutions and was found to be linearly proportional to the model blood flow rate. In an experimental dialysis system using a xanthan gum solution (300 mg/L), the A-V pressure gradient showed a gradual linear increase along with the ultrafiltration rate up to 1 L/h as the viscosity slowly increased in the dialyzer. The changes of blood volume shown by this method were significantly correlated with data obtained using the Crit-Line in 8 patients undergoing hemodialysis. This simple and inexpensive method may allow monitoring of blood volume changes and thus provide data that are beneficial for fluid management in hemodialysis patients suffering from clinical dialysis intolerance.

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure01:10

Pre-Procedural Guidelines for Assessing Blood Pressure

Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the patient.
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...