Related Experiment Video
Updated: Aug 23, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
[Creation and clinical application of real-time dose monitor using dose area product meter]
Kosuke Matsubara1, Kichiro Koshida, Atsushi Fukuda
1Department of Radiology, Kanazawa University Hospital.
Abstract:
The management of patient dose has become more of an issue in recent years. Dose can be determined non-invasively and in real time through the use of a dose area product meter, but it is the area dose value that is obtained. Therefore, we created a program that estimates entrance skin dose (ESD) in real time from area dose values obtained during procedures. We used Microsoft Visual C++ 6.0 (Standard Edition) for the programming language and C language for the programming environment. The value was a maximum 285.4 mGy at ileus tube insertion when measuring ESD for radiography of the digestive organ and non-vascular type IVR using the created program and seeking the average according to the procedures. The program that we created can be considered valid for monitoring ESD correctly and in real time.