Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
THEORETICAL DESIGN CONSIDERATIONS AND PHYSIOLOGIC PERFORMANCE CRITERIA FOR AN IMPROVED INTRACORPOREAL (ABDOMINAL)
Stephen R. Igo1, C Wayne Hibbs, John M. Fuqua
1Cardiovascular Surgical Research Laboratories, Texas Heart Institute of St. Luke's Episcopal and Texas Children's Hospitals, P.O. Box 20269, Houston, Texas 77025.
Abstract:
Our laboratories are engaged in the design of a clinically-oriented electrically actuated long-term intracorporeal (abdominal) left ventricular assist device ("E-type" ALVAD) or partial artificial heart. This infradiaphragmatic blood pump is designed to be powered by implantable electrical to mechanical energy converter systems. THE FOLLOWING ANALYSES WERE UNDERTAKEN TO: [List: see text] The proposed "E-type" ALVAD should be capable of pumping 4-7 liters per minute at heart rates of 75-100 beats per minute during rest, and 10 liters per minute at rates of 120 beats per minute during moderate exercise. These performance levels should be exceeded with a maximum device stroke volume of 85-90 ml and a mean pump inflow (filling) impedance of = 0.072 gm/sec/cm(-5).