Disc erosion in Models 103 and 104 of Beall mitral valve prostheses
Ricardo Gómez1, María José Verduras, Alfonso Lopez-Quintana
1Department of Thoracic and Cardiovascular Surgery, La Paz Hospital. Autonomous University of Madrid, Madrid-34, Spain.
Abstract:
Three cases of severe disc variance and erosion of the Teflon-disc Beall mitral valve prosthesis (Models 103 and 104) are reported. In two patients, the Beall mitral valves were excised and replaced with two Björk-Shiley mitral valves. The remaining patient did not survive, and at autopsy, the lens was found at the aortic bifurcation level. Because of this potentially lethal complication, careful follow-up of patients with Beall mitral valve prostheses (Models 103 and 104) is recommended.