Related Experiment Video
Updated: Aug 2, 2026

Puncture-Induced Iris Neovascularization as a Mouse Model of Rubeosis Iridis
Published on: March 8, 2018
Surgical management of iris defects with prosthetic iris devices
I Mavrikakis1, E Mavrikakis, P P Syam
1Sussex Eye Hospital, Brighton and Sussex University Hospitals, Brighton, UK.
Purpose:
To evaluate the safety and efficacy of surgical implantation of prosthetic iris devices in patients with iris deficiency.
Methods:
Nine patients with traumatic iris defects, congenital aniridia or iris coloboma, and surgical or optical iridectomies were included in a noncomparative case series. Cataract surgery with intraocular lens and prosthetic iris implantation was performed in 10 eyes. The visual acuity, subjective degree of glare disability, postoperative anatomic results, and intraoperative and postoperative complications were evaluated.
Results:
The mean follow-up was 17.75 months (range 4-48 months). Best-corrected visual acuity improved in nine of 10 eyes (90%) and remained unchanged in one eye. Glare subjectively improved in four of five eyes (80%) of patients complaining of glare preoperatively. All eyes achieved the desired anatomic result. Intraoperative complications included one anterior capsular tear. Postoperative complications included a short period of mild postoperative anterior uveitis in four eyes. Secondary glaucoma was absent.
Conclusion:
In patients with iris deficiency, implantation of prosthetic iris device, and intraocular lens implant following cataract surgery appears to be safe and effective in reducing glare disability and improving visual outcomes.