Verbal communication in MR environments: effect of MR system acoustic noise on speech understanding
Adriaan Moelker1, Ronald A J J Maas, Peter M T Pattynama
1Department of Radiology, Erasmus Medical Center Rotterdam, 50 Dr Molewaterplein, PO Box 1738, 3000 DR Rotterdam, the Netherlands. a.moelker@erasmusmc.nl
Purpose:
To assess the masking effect of magnetic resonance (MR)-related acoustic noise and the effect of passive hearing protection on speech understanding.
Materials And Methods:
Acoustic recordings were made at 1.5 T at patient and operator (interventionalist in the MR suite) locations for relevant pulse sequences. In an audiologic laboratory, speech-to-noise ratios (STNRs) were determined, defined as the difference between the absolute sound pressure levels of MR noise and speech. The recorded noise of the MR sequences was played simultaneously with the recorded sentences at various intensities, and 15 healthy volunteers (seven women, eight men; median age, 27 years) repeated these sentences as accurately as possible. The STNR that corresponded with a 50% correct repetition was used as the measure for speech intelligibility. In addition, the effect of passive hearing protection on speech intelligibility was tested by using an earplug model.
Results:
Overall, speech understanding was reduced more at operator than at patient location. Most problematic were fast gradient-recalled-echo train and spiral k-space sequences. As the absolute sound pressure level of these sequences was approximately 100 dB at patient location, the vocal effort needed to attain 50% intelligibility was shouting (>77 dB). At operator location, less effort was required because of the lower sound pressure levels of the MR noise. Fast spoiled gradient-recalled-echo and echo-planar imaging sequences showed relatively favorable results with raised voice at operator location and loud speaking at patient location. The use of hearing protection slightly improved STNR.
Conclusion:
At 1.5 T, the level of MR noise requires that large vocal effort is used, at the operator and especially at the patient location. Depending on the specific MR sequence used, loud speaking or shouting is needed to achieve adequate bidirectional communication with the patient. The wearing of earplugs improves speech intelligibility.
Related Concept Videos
Barriers to Effective Communication I
Communication barriers include the following:
Physiological barriers: They are limitations caused by a person's health condition or disability, such as hearing loss, poor eyesight, illness, or unconsciousness. An example to overcome this barrier...
Barriers to Effective Communication II
Cultural barriers:
Differences in values, beliefs, religion, knowledge, and tradition can significantly impact communication. Awareness of nonverbal cues is critical, especially when conversing with a patient from a different culture. What appears appropriate in one culture may be inappropriate in another.
Semantic barriers:
As a result of their tendency to use...
Imaging Studies for Cardiovascular System IV: CMRI