Methodological aspects of high-frequency ultrasound of skin in children
A J Schou1, K Thomsen, A M Plomgaard
1Children's Clinic Randers, Dytmaersken 9,3. Randers, Denmark. ajs@dadlnet.dk
Background:
High-frequency ultrasound of the skin has recently been introduced for assessment of systemic effects in the cutis and subcutis of oral and inhaled glucocorticoids in children. However, the use of high-frequency skin ultrasound in clinical trials is invalidated because important methodological aspects have not been addressed. The aim of the present study was to evaluate inter- and intraobserver, day to day and diurnal variations of measurement of thickness of cutis and subcutis, and the fraction of low echogenic pixels (fLEP) in the cutis and, furthermore, to assess effects of exercise on the cutis and subcutis and variations in subcutaneous thickness between anatomical locations in children with a high-frequency B-mode ultrasound scanning device.
Methods:
Three studies were conducted, each including 10 healthy prepubertal children. High-frequency skin ultrasound was performed with the 20 MHz Dermascan C (Cortex Technology, Hadsund, Denmark). In study 1, the same observer performed five consecutive scannings to assess intraobserver variations. In study two different observers performed scannings at 2 h intervals between 08:00 and 20:00 h, whereby interobserver and diurnal variations were assessed. In study 3, the same observer performed scannings in different anatomical locations on five consecutive days, and on one of these days before and after exercise. Thus day-to-day variations and the effect of exercise were assessed.
Results:
Low inter- and intraobserver variations were found on assessment of the thickness of cutis and subcutis, whereas high variations were found on evaluation of the dermal water content. Diurnal variations were absent, and day-to-day variations were low. Exercise caused significant increases in the thickness of cutis and subcutis on the thigh.
Conclusion:
Low inter- and intraobserver variations make high-frequency ultrasound a precise and reliable tool for assessment of the cutaneous and subcutaneous thickness in children. In future trials, repetitive scannings need not to be performed at the same time of the day, whereas strenuous physical activity should be avoided on days of examination.
More Related Videos
06:43High Frequency Ultrasound for the Analysis of Fetal and Placental Development In Vivo
Published on: November 8, 2018
09:05Generation and Quantitative Analysis of Pulsed Low Frequency Ultrasound to Determine the Sonic Sensitivity of Untreated and Treated Neoplastic Cells
Published on: July 22, 2015
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: