Use of laser to extract unwanted pacemaker leads
1Department of Cardiovascular Surgery and the Department of Clinical Engineering, Shadyside Hospital, Pittsburgh, Pennsylvania, USA.
Abstract:
Of the total number of pacemakers implanted each year, 6% must eventually be replaced because of infection, lead failure, or other problems. Often however, removal of the pacing leads is precluded by scarring at the tip of the leads. Because abandoned leads can cause serious problems, we initiated the present study to test the feasibility of using a laser to remove such leads. After performing in vitro experiments to establish the laser settings required for severing leads, we implanted 6 leads, from 4 different manufacturers, in 3 dogs. Five weeks later, we passed a laser fiber through each lead to the tip and delivered a 10-watt impulse for 2 seconds, which resulted in the successful removal of all the leads. When the dogs were killed 3 days later, postmortem examination revealed no thromboembolism in the heart or lungs and no myocardial damage. Subsequent experiments in 3 human cadaver hearts-involving leads implanted for 5 years, 2 years, and 2 months-yielded the same satisfactory results as did the canine experiments.