Relevance of type of catheters for central venous pressure measurement
Samuel Santelices1, Kevin J Sullivan, Niranjan Kissoon
1University of Florida, Health Science Center, Jacksonville, FL, USA.
Objective:
To compare simultaneous central venous pressure measurements from rigid polyurethane and soft tunneled silicone elastomere catheters.
Hypothesis:
There will be no significant difference in central venous pressure readings between polyurethane and silastic catheters.
Setting:
Bone Marrow Transplant Unit in a tertiary care children's hospital.
Patients:
Five children undergoing bone marrow transplantation with preexisting polyurethane and silastic catheters.
Methods:
Simultaneous central venous pressure readings were obtained by 2 observers blinded to the other readings and to the type of catheter. Readings were done in triplicate (total of 690 readings). Each triplicate was averaged to 1 data point yielding 115 paired central venous pressure measurements.
Results:
No significant difference was demonstrated between polyurethane and silicone catheters (-1 +/- 3 cm H20). Using Bland and Altman method revealed no significant bias (mean = -1 cm H2O) and acceptable agreement between catheter types.
Conclusion:
Silicone and polyurethane catheters yield similar values of central venous pressures. Permanently implanted silicone elastomere catheters can be used to measure central venous pressure in the emergency setting.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
Related Concept Videos
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Cardiac Catheterization II: Right Heart Catheterization
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Catheterization IV: Nursing Management
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies