A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
A new ultrasound-guided esophageal variceal pressure-measuring device
Larry S Miller1, Qing Dai, Antonia Thomas
1Department of Gastroenterology, Temple University Hospital, Philadelphia, Pennsylvania 19140, USA.
Objective:
To develop a noninvasive method and device to determine intravariceal pressure and variceal wall tension by measuring the variables of the Laplace equation and test this device in a model of esophageal varices.
Methods:
Two variceal pressure measurement devices were constructed. The first device consists of an Olympus 20 MHz ultrasound transducer placed next to a latex balloon catheter attached to a pressure transducer. The second device was constructed by placing the same ultrasound transducer inside a latex condom balloon attached to a pressure transducer. These pressure measurement devices were tested blindly in varix models with different intravariceal pressures, by inflating the balloon to flatten the varix models. Each variceal pressure was measured 10 times by two separate investigators blinded to the actual pressures. The mean intravariceal pressures were calculated. The variceal models were made of a latex balloon filled with water and coffeemate.
Results:
The correlation coefficient between the actual and measured varix pressures for both devices was 0.99. The percent error ranged from 0 to 10%. The correlation coefficient between the investigators making the blinded measurements for both devices was 0.98.
Conclusion:
Two pressure-measuring devices were developed to determine intravariceal pressure in a model varix system. These devices demonstrate a low percent error and a high correlation to the actual variceal pressures with low intra- and interobserver variability. These devices have the potential to measure all the variables of the Laplace equation for wall tension. We plan to test these devices in human subjects.
Related Concept Videos
Esophageal Varices-II: Clinical Features and Management
In the initial assessment, a thorough review of the patient's medical history is vital to identify risk factors such as liver disease, alcohol abuse, or...
Esophageal Varices-I: Introduction
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...