Abstract:
Finger arterial blood pressures determined by a newly developed sphygmomanometer, HEM-802F, were compared with arterial pressure obtained from direct measurement of the radial artery. An excellent correlation was found between the two methods (systolic: r = 0.93, diastolic: r = 0.91), although there was a large variability among individual subjects. The range of differences between them are from +32 to -13 mmHg for systolic and +15 to -25 mmHg for diastolic blood pressure measurement. HEM-802F underestimated systolic pressure (-4.0 mmHg) and overestimated diastolic pressure (+6.7 mmHg), compared with intra-arterial readings. The HEM-802F was useful for the non-invasive arterial pressure monitoring during general anesthesia.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure