X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Application of low dose rate pulsed fluoroscopy in cardiac pacing and electrophysiology: patient dose and image
C J Kotre1, S Charlton, K J Robson
1Regional Medical Physics Department, Newcastle General Hospital, Newcastle-upon-Tyne NE7 7DN, UK.
Abstract:
The performance of a low dose rate pulsed fluoroscopy option and its successful application to cardiac pacing and electrophysiology is reported. Low dose rate 6.25 frames per second pulsed fluoroscopy was made available in two catheter laboratories at a specialist cardiac centre in February 2003, and was adopted as the standard imaging technique for cardiac pacing procedures. The image quality was found to be considerably poorer than conventional modern units, being very similar to that which would have been accepted as adequate performance 20 years ago, but at less than one-tenth of the dose rate. No problems with the clinical acceptance of this imaging mode for cardiac pacing and electrophysiology have been reported. The already low median patient dose-area product for pacing at this cardiac centre was further reduced by 50% with the introduction of this fluoroscopy option.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT