Dose reduction in computed tomography by individualized scan protocols.

L Jangland1, E Sanner, J Persliden

  • 1Department of Medical Physics, University Hospital, Uppsala, Sweden. lars.jangland@akademiska.se

Acta Radiologica (Stockholm, Sweden : 1987)
|July 9, 2004
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors.

The British journal of radiology·2015
Same author

ICRP Publication 117. Radiological protection in fluoroscopically guided procedures performed outside the imaging department.

Annals of the ICRP·2012
Same author

Optimization of computed tomography urography protocol, 1997 to 2008: effects on radiation dose.

Acta radiologica (Stockholm, Sweden : 1987)·2009
Same author

Optimizing the tube potential for lumbar spine radiography with a flat-panel digital detector.

The British journal of radiology·2008
Same author

Massive postpartum hemorrhage treated with transcatheter arterial embolization: technical aspects and long-term effects on fertility and menstrual cycle.

Acta radiologica (Stockholm, Sweden : 1987)·2007
Same author

Impact of digital imaging on radiation doses to the patient during X-ray examination of the urinary tract.

Acta radiologica (Stockholm, Sweden : 1987)·2005

This study introduces methods to adjust computed tomography (CT) radiation dose based on patient size, significantly reducing radiation exposure, especially for smaller individuals.

Area of Science:

  • Medical Imaging
  • Radiology
  • Radiation Physics

Background:

  • Patient radiation dose in computed tomography (CT) is a significant concern.
  • Optimizing radiation dose while maintaining image quality is crucial for patient safety.

Purpose of the Study:

  • To develop and validate methods for adjusting the milliampere-second (mAs) value in CT examinations based on patient size.
  • To minimize radiation dose delivered to patients during CT scans.

Main Methods:

  • Two mathematical methods were employed to calculate patient-specific correction factors for tube charge (mAs).
  • The methods were validated using Perspex phantoms of varying sizes and geometries.
  • Image noise was quantified before and after applying correction factors.
  • Correction factors were retrospectively applied to 12 abdominal CT examinations to estimate dose reduction.

Related Experiment Videos

Main Results:

  • Both mathematical methods significantly reduced image noise variations across different phantom sizes.
  • Retrospective analysis of patient data indicated potential dose reduction factors of up to 7 in extreme cases.
  • The largest calculated correction factor was seven times greater than the smallest.

Conclusions:

  • The proposed methods for adjusting mAs based on patient size are applicable to most CT systems.
  • Significant radiation dose reduction is achievable, particularly for pediatric and smaller adult patients.
  • This approach enhances patient safety in CT imaging.