Rectangular collimation and radiographic efficacy in eight general dental practices in the West Midlands
Related Experiment Video
Updated: Aug 23, 2026

Radiation Planning Assistant - A Web-based Tool to Support High-quality Radiotherapy in Clinics with Limited Resources
Published on: October 6, 2023
Peter H Thornley1, Dominic A Stewardson, P G John Rout
1pthornley@blueyonder.co.uk
Introduction:
Rectangular collimation is recommended in order to reduce radiation doses to patients. However, anecdotal reports suggest that it may result in more retakes, with a possible net increase in patient dosage.
Method:
To test this hypothesis, eight general dental practitioners were recruited to participate in a project. Each took 25 pairs of bitewing radiographs of patients before and 25 after fitting rectangular collimators to their x-ray machines. Double-pack films were used so that duplicates could be examined. These were assessed by a consultant oral radiologist according to the National Radiographic Protection Board (NRPB) grading system for positioning and general film quality.
Results:
Initially, positioning quality was above the target level for six of the eight dentists; however, for all but one, this dropped after using rectangular collimation. General film quality was of a lower overall standard initially. Using rectangular collimation had an effect on general film quality but the direction and size of this varied among the dentists.
Conclusion:
It was concluded that amongst the participating dentists, although numbers of retakes increased by 7%, there would be a net reduction in dosage to patients following rectangular collimation.