Related Experiment Video
Updated: Aug 23, 2026

Biological Compatibility Profile on Biomaterials for Bone Regeneration
Published on: November 16, 2018
[Biocompatibility testing of various biomaterials as dependent on immune status]
S Endres1, M Landgraff, M Kratz
1Klinik für Orthopädie und Rheumatologie der Philipps-Universität Marburg.
Introduction:
This study deals with the ingrowth behaviour of biomaterials (hydroxyapatite, cp-titanium, cobalt-chromium-molybdenum and PAEK) in relationship to the immunological competence in an animal model. Measured were the production of extracellular matrix (ECM) after implantation in non-immunocompetent naked mice and immunocompetent wild mice. Intention of the trial was to find out if either the immunological competence or the duration of implantation influences the quantity of produced ECM. In addition, the ingrowth behaviour was investigated under these conditions by using four different biomaterials.
Material And Methods:
Biomaterials (hydroxyapatite, cp-titanium, cobalt-chromium-molybdenum and PAEK) were implanted for 14 or 60 days, respectively. CLSM, SEM and SEM-EDX were used for analysis of the ECM and for measuring the distance between ECM and the biomaterials. CLSM was also used for the detection of collagen I and III as a parameter of the quality of osteointegration.
Results:
In all cases a matrix grew on the surface of the biomaterials. The CLSM detected a co-localisation of collagen I and III. In the case of hydroxyapatite collagen I and III were found at a distance of 1 micro m over the surface. The largest space between the surface of the implant and the ECM was found in the case of PAEK. The smallest space was in the case of hydroxyapatite. In all investigated biomaterials the proportion of collagen I to collagen III varied through the duration of implantation.
Discussion:
As is known from the literature we found different ingrowth behaviours on using different biomaterials. Furthermore, we found a statistically significant influence of the immunological competence of the host with regard to ECM production. We draw the conclusion that immunological competence improves the ingrowth behaviour of biomaterials.