[A simple method for evaluation of the stability of magnetic resonance scanner during brain function imaging]
Xi-gang Feng1, Zhi-bo Wen, Fan-heng Huang
1Department of Medical Instrumentation, First Military Medical University, Guangzhou 510515 China. fxg@fimmu.com
Objective:
To develop a simple method for evaluating the stability of magnetic resonance (MR) scanner during brain function imaging.
Method:
A symmetrical water phantom was scanned using the same sequence and parameters as for brain function imaging. The stability of MR scanner was evaluated in view of the signal levels of phantom images.
Result:
The quality control chart of the mean value, standard deviation and the range of the signal level in the region of interest (ROI) within the same region on 128 phantom images were obtained, demonstrating the state of stability of the MR scanner.
Conclusion:
The quality control chart of ROI can be used to evaluate the stability of the MR scanner.
More Related Videos
17:16Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies for Cardiovascular System IV: CMRI