The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Evaluation of Frova, single-use intubation introducer, in a manikin. Comparison with Eschmann multiple-use introducer
I Hodzovic1, I P Latto, A R Wilkes
1Department of Anaesthetics and Intensive Care Medicine, University of Wales College of Medicine, Heath Park, Cardiff CF14 4XN, UK. hodzovic@cf.ac.uk
Abstract:
In a randomised cross-over study, 48 anaesthetists attempted to place a Frova single-use introducer, an Eschmann multiple-use introducer and a Portex single-use introducer in the trachea of a manikin set up to simulate a grade 3 laryngoscopic view. The anaesthetists were blinded to success (tracheal placement) or failure (oesophageal placement). Successful placement (proportion, 95% confidence interval) of either the Frova introducer (65%, 50-77%) or the Eschmann introducer (60%, 46-73%) was significantly more likely than with the Portex introducer (8%, 3-20%). There were no significant differences between the success rates for the Frova and the Eschmann introducers. A separate experiment revealed that the peak force exerted by the Frova and Portex introducers was two to three times greater than that which could be exerted by the Eschmann introducer, p < 0.0001, indicating that the single-use introducers are more likely to cause tissue trauma during placement.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques