A continuous-flow instrument for the determination of linear polyacrylamide stability
Karel Klepárník1, Pavel Mikuska
1Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Brno, Czech Republic. klep@iach.cz
Abstract:
Alkaline hydrolysis of high-molecular-mass (10 MDa) linear polyacrylamide, a widely used replaceable sieving medium for the capillary electrophoresis of DNA fragments and proteins, was investigated. The release rate of ammonia, a product of amide group hydrolysis, was monitored by a high-sensitivity continuous-flow system. The experimental results show a rapid onset of hydrolysis at a temperature of 70 degrees C. While the degree of 1 h hydrolysis was evaluated to reach about 16% at 70 degrees C and pH 13, it drops to 5% at 50 degrees C and pH 12.5.
Related Concept Videos
Determination of Molar Masses of Polymers II
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...